Name: | Ati Aerosol Photometer For Cleanroom HVAC System | Reagent Settings: | PAO-4, DOP, Ondina, Mineral Oil, Paraffin |
---|---|---|---|
Output: | Wireless Printer, USB Interface | Light Source: | Solid State, Rated For Life Of Instrument |
Usage: | Leakage Detection Of Clean Room | Flow Control: | Pulse Width Modulation (PWM) Control Ensures Sampling Rate Of 1cfm (28.3lpm) +/- 10% In All Applications. |
High Light: | HVAC System Ati Aerosol Photometer,Cleanroom Ati Aerosol Photometer,Digital Aerosol Photometer For HVAC System |
Ati Aerosol Photometer For Cleanroom HVAC System
The ati Aerosol photometers fpr cleanroom hvac system are primarily used to check for small pinholes and other damage in filter media, such as frame seals, gasket seals, and leaks in the filter frame. By checking the tightness of the high-efficiency filter and its connection with the installation frame, find the defects of the high-efficiency filter itself and its installation in time, and take corresponding remedial measures to ensure the cleanliness of the area.
Features and Benefits of aerosol photometer
Application of digital aerosol photometer
APM-18 photometer Specifications | |
Dynamic Range | Up to 600 micrograms per liter |
Reagent Settings | PAO-4, DOP, Ondina, Mineral Oil, Paraffin |
Auto Zero | Automatically establishes zero reading at startup |
Sensitivity | 1% of readings ≥0.01% to 100% |
Repeatability | 0.5% of readings ≥0.01% to 100% |
Flow Control | Pulse Width Modulation (PWM) control ensures sampling rate of 1cfm (28.3lpm) +/- 10% in all applications. |
Light Source | Solid State, rated for life of instrument |
Reporting | Continuous, Summary, and Monitoring modes |
Alarm | User selectable audible, visual, and vibratory alarms notify the user when the user defined set point is exceeded. |
Size | 260 (W)X160(H)X360(D) |
Weight | 10Kg |
Output | Wireless printer, USB interface |
Power | 220VAC, 50/60HZ |
Standards compliance | NSF 49, IEST, ISO-14644 |