Basic Information
Place of Origin: China
Brand Name: SUJING
Model Number: APM-18
Minimum Order Quantity: 1
Price: Negotiable
Packaging Details: Carton
Delivery Time: 5 work days
Payment Terms: T/T
Supply Ability: 800/per month
Detail Information
Name: APM-18 Detect Leak Test Equipment Didital Aerosol Photometer Repeatability: 0.5% Of Readings ≥0.01% To 100%
Sensitivity: 1% Of Readings ≥0.01% To 100% Flow Control: Pulse Width Modulation (PWM) Control Ensures Sampling Rate Of 1cfm (28.3lpm) +/- 10% In All Applications.
Auto Zero: Automatically Establishes Zero Reading At Startup Dynamic Range: Up To 600 Micrograms Per Liter
High Light:

Detect Leak Test Digital Aerosol Photometer

,

PWM Control Digital Aerosol Photometer

,

Auto Zero Digital Aerosol Photometer


Product Description

APM-18 Detect Leak Test Equipment Didital Aerosol Photometer

Aerosol photometer is designed according to Mie scattering theory. It can be used to detect leakage of highly efficient filters. The device can realize the detection of aerosol upstream and downstream. It can be applied to the clean room, laminar flow, biological safety cabinet, HEPA system, HVAC system, HEPA filter, vacuum filter, and the filter.

 

Features and Benefits of aerosol photometer

  • long life laser light source;
  •  high precision photoelectric multiplier tube detection;
  •  can set PAO, DOP and other types of aerosols;
  • equipped with a dedicated handheld device to achieve control, display and sampling function;
  •  large capacity data storage, real-time data storage;
  • more than the setting alarm sound and light alarm;
  • can be obtained by U disk or thermal printer to print historical data;
  •  can print leakage rate and other monitoring data in real time;
  •  through the special software, the sampling data can be imported into PC;
  •  automatic fault detection.

Application of digital aerosol photometer 

  • leakage detection of clean room
  • VLF bench
  • biosafety cabinet
  • glove box
  • HEPA vacuum cleaner
  • HVAC system
  • HEPA filter
  • negative pressure filtering system
  • operating theatre
  • nuclear filter system
  • collection protect filter
APM-18 photometer Specifications
Dynamic Range Up to 600 micrograms per liter
Reagent Settings PAO-4, DOP, Ondina, Mineral Oil, Paraffin
Auto Zero Automatically establishes zero reading at startup
Sensitivity 1% of readings ≥0.01% to 100%
Repeatability 0.5% of readings ≥0.01% to 100%
Flow Control Pulse Width Modulation (PWM) control ensures sampling rate of 1cfm (28.3lpm) +/- 10% in all applications.
Light Source Solid State, rated for life of instrument
Reporting Continuous, Summary, and Monitoring modes
Alarm User selectable audible, visual, and vibratory alarms notify the user when the user defined set point is exceeded.
Size 260 (W)X160(H)X360(D)
Weight 10Kg
Output Wireless printer, USB interface
Power 220VAC, 50/60HZ
Standards compliance NSF 49, IEST, ISO-14644

 

 